Hagemeier E, Kemper K, Boos K S, Schlimme E
J Clin Chem Clin Biochem. 1984 Feb;22(2):175-84. doi: 10.1515/cclm.1984.22.2.175.
We describe an on-line multi-column high performance liquid chromatographic method for the selective clean-up and analysis of major and minor ribonucleosides in physiological fluids. Quantitative data obtained for the determination of some methylated ribonucleosides in human urines are compared with those obtained with the traditional off-line method. The on-line technique developed in our laboratory is distinguished from the off-line method by the following features: Sample clean-up and analysis of the target-compounds can easily be automatized, Total time of analysis, for example of urinary ribonucleosides, is decreased to 35 minutes, Laborious and error-prone evaporation and redissolution steps are avoided, Reliability of the overall analytical system can be controlled with ease, Small sample-volumes can be applied directly, Sensitive samples can be processed very rapidly under mild conditions, Results obtained with the on-line and off-line-techniques compare well.