Etowski D C, Beckwith D G
Am J Clin Pathol. 1981 Jun;75(6):830-3. doi: 10.1093/ajcp/75.6.830.
The accuracy of quantitative aminoglycoside minimal inhibitory concentration (MIC) determinations was evaluated with the Micro-Media Systems microdilution MIC panel and Pseudomonas aeruginosa. Without Mg++ and Ca++ supplementation, very major interpretive discrepancies occurred. A simple, in-expensive cation supplementation procedure was evaluated to correct the discrepancies. The necessity of cation supplementation for susceptibility testing of P. aeruginosa when employing MIC procedures is stressed.
使用微媒体系统微量稀释MIC板和铜绿假单胞菌评估了定量氨基糖苷类最低抑菌浓度(MIC)测定的准确性。在不补充Mg++和Ca++的情况下,出现了非常大的解释性差异。评估了一种简单、廉价的阳离子补充程序以纠正这些差异。强调了在采用MIC程序时对铜绿假单胞菌进行药敏试验补充阳离子的必要性。