Patel K F, Hicks J N
Anesth Analg. 1981 Dec;60(12):885-8.
In vitro experiments were performed to test the flammability of endotracheal tubes in the presence of CO2 laser beams Two types of tubes (Rusch and Portex) were exposed to laser beams at 15 or 30 W intensity for varying periods of time oxygen environments of 25% or 60%. Red rubber Rusch tubes were significantly (p less than 0.01) more resistant to ignition than plastic Portex tubes at both higher energy exposures and greater durations of exposure. Aluminum tape wrapping significantly (p less than 0.01) retarded ignition of both types of tubes. In view of this, a clinical protocol was developed using aluminum tape-wrapped red rubber tubes in 98 patients for 189 procedures requiring the use of a laser. In no case was ignition encountered.
进行了体外实验,以测试在二氧化碳激光束存在的情况下气管导管的可燃性。两种类型的导管(Rusch和Portex)在15瓦或30瓦强度的激光束下,于25%或60%的氧气环境中暴露不同时间。在更高能量暴露和更长暴露持续时间下,红色橡胶Rusch导管比塑料Portex导管对点火的抵抗力显著更强(p小于0.01)。铝带包裹显著(p小于0.01)延迟了两种导管的点火。鉴于此,制定了一项临床方案,在98例患者的189次需要使用激光的手术中使用铝带包裹的红色橡胶导管。未发生一例点火情况。