Al-Ghazi M S, Lingman D, Arjune B, Gilbert L D, Thekkumthala J
Department of Medical Physics, Thunder Bay Regional Cancer Centre, Ontario Cancer Treatment and Research Foundation, Canada.
Med Phys. 1991 Jul-Aug;18(4):821-8. doi: 10.1118/1.596637.
Dosimetry measurements have been carried out for the electron beams produced by a linear accelerator at energies 6, 8, 10, 14, 18, and 21 MeV. Characteristic parameters of the central axis dose distributions were derived and compared to corresponding values of electron beams from other accelerators in clinical use where such a comparison is appropriate. A comprehensive set of dosimetric parameters is provided for electron beam treatment planning. The data include central axis depth dose, range-energy parameters, beam penumbra and uniformity.
已对一台直线加速器在6、8、10、14、18和21兆电子伏特能量下产生的电子束进行了剂量测定。推导了中心轴剂量分布的特征参数,并与其他临床使用的加速器产生的电子束的相应值进行了比较(在合适的情况下进行这种比较)。为电子束治疗计划提供了一套全面的剂量测定参数。数据包括中心轴深度剂量、射程 - 能量参数、射束半值层和均匀性。