Gaydos J M, Harrington B J
Antimicrob Agents Chemother. 1982 Mar;21(3):516-8. doi: 10.1128/AAC.21.3.516.
The standard disk diffusion method of Bauer et al. (Am. J. Clin. Pathol. 45:493-496, 1966; National Committee for Clinical Laboratory Standards, Performance standards for antimicrobial disc susceptibility tests, approved standard ASM-2, 2nd ed., 1979) was used to monitor the potency of 16 Autobac antibiotic elution disks over an 8-month period. Results obtained by using the three recommended Autobac control strains showed excellent reproducibility. Zones sizes were all within a 5-mm range, and standard deviations ranged from 0.6 to 1.3 mm.
采用鲍尔等人的标准纸片扩散法(《美国临床病理学杂志》45:493 - 496, 1966;美国国家临床实验室标准委员会,抗菌药物纸片药敏试验性能标准,批准标准ASM - 2,第2版,1979),在8个月的时间内监测16种自动细菌鉴定仪抗生素洗脱纸片的效能。使用三种推荐的自动细菌鉴定仪对照菌株获得的结果显示出极佳的重现性。抑菌圈大小均在5毫米范围内,标准差为0.6至1.3毫米。