Oh I, Carlson C E, Tomford W W, Harris W H
J Bone Joint Surg Am. 1978 Jul;60(5):608-13.
When the distal part of the medullary canal of the femur was plugged with a bolus of methylmethacrylate prior to the insertion of cement and femoral component, fixation of the prosthesis was improved in vitro. A special syringe was devised to introduce the plug at the desired level. When the plug was used, the penetration of the cement into the trabecular bone lining the canal was increased, and subsequent push-out tests showed that the force necessary to disrupt the methacrylate from the bone was significantly greater. Casts of the methacrylate from femora with and without plugged canals showed that the surfaces of the casts made with the canal plugged conformed much more completely with the irregularities of the bone and provided more intimate contact at the cement-bone interface.
在植入骨水泥和股骨假体之前,用一团甲基丙烯酸甲酯堵塞股骨骨髓腔的远端部分,可在体外改善假体的固定。设计了一种特殊的注射器,以便在所需水平插入堵塞物。使用堵塞物时,骨水泥向髓腔内壁小梁骨的渗透增加,随后的推出试验表明,将甲基丙烯酸甲酯从骨中分离所需的力显著更大。对有堵塞和无堵塞髓腔的股骨的甲基丙烯酸甲酯铸型进行观察,结果显示,堵塞髓腔制作的铸型表面与骨的不规则形状更完全地贴合,在骨水泥-骨界面提供了更紧密的接触。