Blijenberg B G, Brouwer H J, Leijnse B
J Clin Chem Clin Biochem. 1978 Aug;16(8):425-8. doi: 10.1515/cclm.1978.16.8.425.
The interference of hemoglobin in the determination of total bilirubin with the Du Pont ACA was studied. The method seemed of limited value in our opinion especially with respect to neonatal sera. By changing the wave-length setting a rather broad plateau was found when the hemoglobin concentration was plotted against the actual bilirubin concentration. On the basis of these findings a new calibration graph was constructed and programmed into the ACA. The new method correlates well with two other micro bilirubin techniques.
研究了血红蛋白对使用杜邦自动临床分析仪(Du Pont ACA)测定总胆红素的干扰。我们认为该方法的价值有限,尤其是对于新生儿血清。通过改变波长设置,当将血红蛋白浓度与实际胆红素浓度作图时,发现了一个相当宽的平稳段。基于这些发现,构建了一条新的校准曲线并编入了自动临床分析仪中。新方法与另外两种微量胆红素检测技术的相关性良好。