The isolation of lipoprotein fractions in newborn plasma was performed by ultracentrifugation in an Air-driven ultracentrifuge (Airfuge). The purity and recovery of the fractions was checked by quantitation of the apo A1 and B proteins in the supernatant and infranatant fractions, and by gradient gel electrophoresis in 4-30% polyacrylamide gels. The influence of the duration of the ultracentrifugal run, the temperature and method for recovery of the fractions were tested. Under optimal conditions pure VLDL fractions could be isolated at d = 1.006 milligrams, whereas VLDL + LDL isolated at d = 1.063 milligrams were contaminated with HDL and other plasma proteins. Isolation of total lipoproteins at d = 1.21 milligrams enabled recovery of 75% of total plasma lipoproteins, however strongly contaminated by other plasma proteins. These results indicate that the recovery and purity of the lipoprotein fractions isolated in the Airfuge are not comparable to the results obtained in a conventional preparative ultracentrifuge.
采用空气驱动超速离心机(Airfuge)通过超速离心法分离新生儿血浆中的脂蛋白组分。通过对上清液和沉淀组分中的载脂蛋白A1和B蛋白进行定量分析,以及在4 - 30%聚丙烯酰胺凝胶中进行梯度凝胶电泳,来检测各组分的纯度和回收率。测试了超速离心运行时间、温度和组分回收方法对分离效果的影响。在最佳条件下,可在密度d = 1.006毫克时分离出纯的极低密度脂蛋白(VLDL)组分,而在密度d = 1.063毫克时分离得到的VLDL +低密度脂蛋白(LDL)组分被高密度脂蛋白(HDL)和其他血浆蛋白污染。在密度d = 1.21毫克时分离总脂蛋白,可回收75%的血浆总脂蛋白,但受到其他血浆蛋白的严重污染。这些结果表明,在Airfuge中分离得到的脂蛋白组分的回收率和纯度与传统制备型超速离心机所获得的结果不可比。