Severin E, Ohnemus B, Kiegler S
Cytometry. 1983 Jan;3(4):308-10. doi: 10.1002/cyto.990030415.
A flow cytometer of the impulse cytophotometer (ICP)-type has been developed for two-parameter flow cytometry. It uses two different methods of excitation simultaneously: an argon laser beam focused onto the cell stream within a modified flow chamber, and the light of a high pressure mercury (Hg) lamp using Köhler illumination. The two fluorescence signals are collected sequentially by the same photomultiplier tube. The signals are processed on a cell-by-cell basis using new hardwired analog signal processing electronics enabling their accumulation and display as two-parameter frequency distribution histograms.
一种用于双参数流式细胞术的脉冲细胞光度计(ICP)型流式细胞仪已被开发出来。它同时使用两种不同的激发方法:一种是聚焦在改进型流动室内细胞流上的氩激光束,另一种是使用柯勒照明的高压汞(Hg)灯发出的光。两个荧光信号由同一个光电倍增管顺序收集。信号通过新的硬连线模拟信号处理电子设备逐细胞进行处理,从而能够将它们累积并显示为双参数频率分布直方图。