通过热稀释技术测量射血分数。

Measurement of ejection fraction by thermal dilution techniques.

作者信息

Kay H R, Afshari M, Barash P, Webler W, Iskandrian A, Bemis C, Hakki A H, Mundth E D

出版信息

J Surg Res. 1983 Apr;34(4):337-46. doi: 10.1016/0022-4804(83)90081-1.

Abstract

The reproducibility, accuracy, and clinical applicability of ventricular ejection fraction derived by a thermal dilution technique were assessed in 22 dogs and 18 patients. Results obtained by the thermal technique were compared to simultaneous results obtained by radionuclide angiography. Right ventricular ejection fraction, measured in 9 dogs (1014 determinations) and 8 patients (744 determinations) was reproducible +/- 5%. Left ventricular ejection fraction, measured in 10 patients, was reproducible +/- 5%. Correlation between thermal and radionuclear measurements varied from 0.86 to 0.93 (all P less than 0.02). We conclude that, because of its low cost, ease of use, and accuracy, thermally derived ejection fraction determinations can be helpful in hemodynamic monitoring of critically ill patients.

摘要

采用热稀释技术测定心室射血分数的可重复性、准确性及临床适用性,对22只犬和18例患者进行了评估。将热稀释技术获得的结果与放射性核素血管造影同时获得的结果进行比较。在9只犬(1014次测定)和8例患者(744次测定)中测得的右心室射血分数的可重复性为±5%。在10例患者中测得的左心室射血分数的可重复性为±5%。热稀释法与放射性核素测量法之间的相关性在0.86至0.93之间(所有P均小于0.02)。我们得出结论,由于热稀释法测定射血分数成本低、使用方便且准确性高,因此有助于对危重症患者进行血流动力学监测。