Anderson R M, Livnat A, Zehr J E
Am J Physiol. 1983 Jan;244(1):H162-5. doi: 10.1152/ajpheart.1983.244.1.H162.
A crystal-controlled telemetry system was developed for long-term high-reliability recording of blood pressure in unrestrained animals. The miniaturized transducer-transmitter system is powered by two lithium batteries that allow for more than 200 h of continuous recording of blood pressure before battery replacement. The total system has a flat frequency response of 0-25 Hz and a time stability of +/- 1 mmHg/day. Linearity is better than 0.5%. The high degree of dependability of the system permits us to perform continuous, unattended, around-the-clock recordings of pressure and heart rate in unrestrained animals. Very high dependability is achieved by the use of a novel piezoresistive blood pressure transducer and by the inherent stability of the crystal-controlled radio-frequency oscillator of the transmitter. The receiver is also crystal tuned to the measured transmitter frequency. This transmitter-receiver frequency locking ensures very reliable reception of the telemetered signal.
开发了一种晶体控制遥测系统,用于在无束缚动物中进行长期高可靠性的血压记录。小型化的换能器-发射器系统由两节锂电池供电,在更换电池前可连续记录血压超过200小时。整个系统的平坦频率响应为0-25Hz,时间稳定性为±1mmHg/天。线性度优于0.5%。该系统的高度可靠性使我们能够在无束缚动物中进行连续、无人值守的全天候压力和心率记录。通过使用新型压阻式血压换能器以及发射器晶体控制射频振荡器的固有稳定性,实现了非常高的可靠性。接收器也通过晶体调谐到所测发射器频率。这种发射器-接收器频率锁定确保了遥测信号的非常可靠的接收。