Suppr超能文献

比较了三种血清铁蛋白的商业检测方法并消除了高剂量“钩状效应”。

Three commercial methods for serum ferritin compared and the high-dose "hook effect" eliminated.

作者信息

Ng R H, Brown B A, Valdes R

出版信息

Clin Chem. 1983 Jun;29(6):1109-13.

PMID:6851105
Abstract

We evaluated four commercial kits for measuring serum ferritin, based on three techniques: immunoradiometric assay, radioimmunoassay, and enzyme immunoassay. The kits evaluated were those manufactured by Abbott Laboratories, Clinical Assays, Corning Medical, and Ramco. Two of the immunoradiometric kits showed satisfactory results with respect to sensitivity and precision; they should be useful in diagnosing individuals with uncomplicated iron-deficiency anemia. One of the immunoradiometric assay kits, however, showed a high-dose "hook effect," beginning at 10 mg of ferritin per liter. We modified this kit to eliminate this effect, at least to ferritin concentrations of 33 mg/L. (We observed a ferritin value as high as 47 mg/L in one patient.) Results with all these kits did not inter-compare well for ferritin concentrations greater than 300 micrograms/L, a finding that casts further doubt on the controversial use of serum ferritin measurement in cases of iron overload.

摘要

我们基于免疫放射分析、放射免疫分析和酶免疫分析这三种技术,对四种用于测定血清铁蛋白的商用试剂盒进行了评估。所评估的试剂盒分别由雅培实验室、临床检验公司、康宁医疗公司和拉姆科公司生产。其中两种免疫放射分析试剂盒在灵敏度和精密度方面显示出令人满意的结果;它们在诊断单纯缺铁性贫血患者时应该会很有用。然而,其中一种免疫放射分析试剂盒在铁蛋白浓度达到每升10毫克时出现了高剂量“钩状效应”。我们对该试剂盒进行了改进,以消除这种效应,至少在铁蛋白浓度达到33毫克/升时能够消除。(我们在一名患者中观察到铁蛋白值高达47毫克/升。)对于铁蛋白浓度大于300微克/升的情况,所有这些试剂盒的结果之间的可比性不佳,这一发现进一步质疑了在铁过载病例中使用血清铁蛋白测量这一存在争议的做法。