Dunn C D, Boden D J
Clin Chem. 1981 Jul;27(7):1280-3.
We evaluated three commercially available immunoradiometric assays for serum ferritin, with particular emphasis on the statistical validity of the results. The data show that it was unusual (a) for dose/response relationships of the standards to be linear over the whole concentration range suggested by the manufacturers and (b) for dose/response relationships of test sera to parallel those of the standards. These findings cast doubt on the ability of any of the assays reliably to discriminate small differences in ferritin concentrations. Nevertheless, all three methods gave reproducible results, but this reflects technical expertise rather than accuracy of the results per se. The method based on liver ferritin as tracer detected less of the iron-binding protein than those based on spleen ferritin, despite significant cross reactivity of the two antibodies.
我们评估了三种市售的血清铁蛋白免疫放射分析方法,特别强调了结果的统计有效性。数据显示,标准品的剂量/反应关系在制造商建议的整个浓度范围内呈线性是不常见的,而且测试血清的剂量/反应关系与标准品的剂量/反应关系平行也是不常见的。这些发现使人怀疑任何一种分析方法能否可靠地区分铁蛋白浓度的微小差异。然而,所有三种方法都给出了可重复的结果,但这反映的是技术专长,而非结果本身的准确性。尽管两种抗体有显著的交叉反应,但基于肝铁蛋白作为示踪剂的方法比基于脾铁蛋白的方法检测到的铁结合蛋白更少。