[不同方法测定封闭型铯-137后装源能量剂量分布的比较]

[Comparison of different methods for the determination of the energy dose distribution of enclosed Cs-137 afterloading sources].

作者信息

Krispel F

出版信息

Strahlentherapie. 1983 May;159(5):306-14.

PMID:6857746
Abstract

A low-dose afterloading unit of the type "Curitron" with Cs-137 sources is available at the Radiologic Hospital of the Karl-Franzens-Universität of Graz. The available sources are between 4.2 cm and 9.1 cm long. The corresponding activities of the individual sources are at present between 117 mCi and 383 mCi. The energy dose distributions around the sources were measured by means of an automatic dose measuring equipment and an additionally developed water phantom. Based on the measurement results, three rapid algorithms were developed and introduced as program modules into the existing therapy planning system which is running under RSX in a minicomputer of the type PDP 11/34. These algorithms are compared with regard to their precision and calculating speed, and the results are presented in form of energy dose distributions.

摘要

格拉茨卡尔 - 弗朗茨大学放射医院有一台配备Cs - 137源的“Curitron”型低剂量后装设备。可用源的长度在4.2厘米至9.1厘米之间。目前各个源的相应活度在117毫居里至383毫居里之间。通过自动剂量测量设备和另外研制的水模体测量源周围的能量剂量分布。基于测量结果,开发了三种快速算法,并作为程序模块引入到现有的治疗计划系统中,该系统在一台PDP 11/34型小型计算机上以RSX运行。对这些算法的精度和计算速度进行了比较,并以能量剂量分布的形式给出了结果。