Isenberg S J, Shoemaker W C
Am J Ophthalmol. 1983 Jun;95(6):803-6. doi: 10.1016/0002-9394(83)90069-7.
The transconjunctival oxygen monitor, mounted on a polymethylmethacrylate conformer, continuously measures the tissue oxygen tension of the palpebral conjunctiva. It is preferable to transcutaneous monitors because it requires no heat and correlates arterial oxygen in cerebral blood flow by monitoring a tissue bed vascularized by the internal carotid artery. We used it on 19 adult patients. The monitor was valuable in correlating arterial oxygen in hemodynamically compromised and uncompromised patients. However, the measurements decreased abruptly when the conjunctival temperature fell below 33 C in the compromised patients. The only ocular complications were conjunctival chemosis and mild punctate keratopathy in some patients; these resolved within 48 hours. This device may be useful in monitoring tissue oxygen tensions in critically ill patients and neonates at risk for retinopathy.
经结膜氧监测仪安装在聚甲基丙烯酸甲酯贴合器上,可连续测量睑结膜的组织氧张力。它优于经皮监测仪,因为它无需加热,通过监测由颈内动脉供血的组织床来关联脑血流中的动脉氧。我们在19名成年患者身上使用了它。该监测仪在关联血流动力学受损和未受损患者的动脉氧方面很有价值。然而,在受损患者中,当结膜温度降至33℃以下时,测量值会突然下降。唯一的眼部并发症是部分患者出现结膜水肿和轻度点状角膜病变;这些症状在48小时内消退。该设备可能有助于监测重症患者和有视网膜病变风险的新生儿的组织氧张力。