[一种抗生素敏感性快速检测技术的评估:使用雅培MS-2自动分析仪常规估算的最低抑菌浓度的可重复性]

[Evaluation of a rapid technic for antibiotic sensitivity testing: reproducibility of minimal inhibitory concentrations routinely estimated with the use of Abbott MS-2 autoanalyzer].

作者信息

Flandrois J P, Coullioud D, Chomarat M, Carret G

出版信息

Ann Biol Clin (Paris). 1983;41(1):55-60.

PMID:6869943
Abstract

The inter-system reproducibility of the MIC results obtained from the MS-2 automated system of antibiotic sensitivity testing was assessed by a two-stage study. In the first stage, a total of 168 MICs were calculated by the automated system from 134 diverse bacterial strains, selected from routine antibiotic sensitivity testing for their "intermediate" sensitivity to one or more antibiotics. The antibiotic sensitivities were repeated using the same machine, in the same laboratory, on another suspension of the same bacteria. The results of antibiotic sensitivity reported as "intermediate" on the first test were compared to those of the second test; only 19 p. cent of the responses remained "intermediate" and the MICs were comparable between the two tests; 50 p. cent became sensitive and 31 p. cent became resistant (MIC not calculated). In the second stage, 4 strains with stable intermediate sensitivity to 8 antibiotics, as shown by 2 routine sensitivity tests, were tested 33, 48, 44, and 39 times respectively against these antibiotics using various suspensions of the same strain. 394 results were obtained: 92 intermediate results with a MIC (24.4 p. cent), 219 sensitive results (56.3 p. cent) and 83 resistant results (20.2 p. cent). The range of values for the MIC calculated by the automatic system was small, so the MICs are reproducible; but, for 5 of the 8 antibiotics, the responses were both "S" and "R". This suggests that the so-called "Intermediate" zone is in fact a zone of uncertainty where the majority of the MIC results are erroneous.

摘要

通过一项两阶段研究评估了从MS - 2抗生素敏感性自动化检测系统获得的最低抑菌浓度(MIC)结果的系统间再现性。在第一阶段,自动化系统从134种不同细菌菌株中计算出总共168个MIC,这些菌株是从常规抗生素敏感性检测中挑选出来的,因其对一种或多种抗生素具有“中介”敏感性。在同一实验室使用同一台机器,对相同细菌的另一份菌悬液重复进行抗生素敏感性检测。将第一次检测报告为“中介”的抗生素敏感性结果与第二次检测结果进行比较;只有19%的结果仍为“中介”,且两次检测的MIC具有可比性;50%变为敏感,31%变为耐药(未计算MIC)。在第二阶段,对4种经两次常规敏感性检测显示对8种抗生素具有稳定中介敏感性的菌株,分别使用同一菌株的不同菌悬液对这些抗生素进行33次、48次、44次和39次检测。共获得394个结果:92个中介结果(MIC)(24.4%),219个敏感结果(56.3%)和83个耐药结果(20.2%)。自动系统计算出的MIC值范围较小,因此MIC具有再现性;但是,对于8种抗生素中的5种,结果既有“敏感”(S)又有“耐药”(R)。这表明所谓的“中介”区域实际上是一个不确定区域,其中大多数MIC结果是错误的。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验