Hangartner T N, Overton T R
J Comput Assist Tomogr. 1982 Dec;6(6):1156-62. doi: 10.1097/00004728-198212000-00017.
A special purpose gamma-ray computed tomography scanner has been developed for precise measurements of bone density in the human appendicular skeleton. Details of the scanner's hardware and of the software organization for system control and data analysis are given, together with an outline of the theoretical basis for conversion of measured linear attenuation coefficients to physical bone densities. Performance of the system was evaluated on bone-like phantoms. Clinically, a precision of +/- 0.5% is obtained for bone density determinations. This device is being used in experimental studies and clinical investigations.
已开发出一种专用伽马射线计算机断层扫描扫描仪,用于精确测量人体附属骨骼的骨密度。文中给出了扫描仪硬件以及用于系统控制和数据分析的软件组织的详细信息,同时还概述了将测量的线性衰减系数转换为物理骨密度的理论基础。该系统在类骨模型上进行了性能评估。在临床上,骨密度测定的精度为±0.5%。该设备正用于实验研究和临床调查。