Lloyd B, Burrin J, Smythe P, Alberti K G
Clin Chem. 1978 Oct;24(10):1724-9.
We describe enzymic fluorometric methods of automated analysis for glucose, lactate, pyruvate, 3-hydroxybutyrate, glycerol, and alanine in perchloric acid extracts of blood. Unmodified Technicon AutoAnalyzer II apparatus is used. The usual concentrations of all these metabolites can be measured in as little as 0.1 ml of blood from a fasting subject. Within-batch and between-batch coefficients of variation ranged from 0.4 to 4.4% for all metabolites except 3-hydroxybutyrate, for which CV's were higher for low concentrations. Analytical recovery of added metabolites ranged from 92 to 98%. Glucose, lactate, alanine, and 3-hydroxybutyrate are stable in perchloric acid extracts for at least 13 days at room temperature, and a year at -20 degrees C; pyruvate shows a 6--8% loss after 3 days and 52% by one year at -20 degrees C; glycerol concentrations were stable at -20 degrees C for at least 13 days. Blank fluorescence is found in perchloric acid extracts of blood, necessitating blank runs for pyruvate, 3-hydroxybutyrate, glycerol, and alanine. The systems are simple to use, relatively inexpensive to operate, and are recommended for any laboratory with high throughput of samples.
我们描述了用于血液高氯酸提取物中葡萄糖、乳酸、丙酮酸、3-羟基丁酸、甘油和丙氨酸自动化分析的酶荧光法。使用未改良的Technicon AutoAnalyzer II仪器。所有这些代谢物的常规浓度可在空腹受试者仅0.1 ml血液中进行测量。除3-羟基丁酸外,所有代谢物的批内和批间变异系数在0.4%至4.4%之间,3-羟基丁酸在低浓度时变异系数较高。添加代谢物的分析回收率在92%至98%之间。葡萄糖、乳酸、丙氨酸和3-羟基丁酸在高氯酸提取物中于室温下至少稳定13天,在-20℃下稳定一年;丙酮酸在3天后损失6%-8%,在-20℃下一年损失52%;甘油浓度在-20℃下至少稳定13天。血液高氯酸提取物中存在空白荧光,因此丙酮酸、3-羟基丁酸、甘油和丙氨酸需要进行空白测定。该系统使用简单,操作成本相对较低,推荐给任何样品通量高的实验室。