Liquid chromatography and fluorescence spectroscopy compared with a homogeneous enzyme immunoassay technique for determining quinidine in serum.

作者信息

Drayer D E, Lorenzo B, Reidenberg M M

出版信息

Clin Chem. 1981 Feb;27(2):308-10.

PMID:7006859
Abstract

Sera from 67 cardiac patients being treated with quinidine were analyzed for this drug by both a homogeneous enzyme immunoassay (EMIT) and a fluorescence procedure (double-extraction method). Of these samples, 47 were also analyzed for quinidine and its major metabolite in serum, (3s)-3-hydroxyquinidine, by liquid chromatography. Results by the EMIT procedure correlated well with those by both the chromatographic and fluorescence procedures, but EMIT gave, on the average, 25% higher values for quinidine than did the chromatographic procedure. The EMIT procedure is therefore not totally specific. (3s)-3-Hydroxyquinidine and dihydroquinidine (a contaminant present in standard quinidine formulations) each show cross reactivity in the EMIT assay for quinidine. Nevertheless, the EMIT assay for quinidine is acceptable for clinical use because these two compounds are also pharmacologically active.

摘要

相似文献

5
A high-pressure liquid chromatography method for serum quinidine and (3S)-3-hydroxyquinidine.
Clin Chim Acta. 1979 Jan 1;91(1):7-13. doi: 10.1016/0009-8981(79)90465-0.
6
Comparison of assay procedures used to measure total and unbound concentrations of quinidine.
DICP. 1989 Dec;23(12):999-1004. doi: 10.1177/106002808902301210.
7
Simple and selective high-performance liquid chromatographic method for estimating plasma quinidine levels.
J Chromatogr. 1980 Feb 8;181(2):207-17. doi: 10.1016/s0378-4347(00)81606-0.

引用本文的文献

1
Free drug concentration monitoring in clinical practice. Rationale and current status.
Clin Pharmacokinet. 1986 Nov-Dec;11(6):450-69. doi: 10.2165/00003088-198611060-00003.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验