Standardization of computed tomography images by means of a material-selective beam hardening correction.

作者信息

Rüegsegger P, Hangartner T, Keller H U, Hinderling T

出版信息

J Comput Assist Tomogr. 1978 Apr;2(2):184-8. doi: 10.1097/00004728-197804000-00012.

Abstract

Polychromaticity of the X-rays used in computed tomography (CT) has made it difficult to establish an absolute scale for CT values and has made quantitative comparisons between patients unreliable. The spectral shift of the X-rays depends on the material distribution within the structure measured and is significant if substantial amounts of bone, fat, or injected contrast material are present. A material-selective beam hardening correction procedure has been developed that allows the reconstruction of good approximations of linear attenuation coefficients with respect to a reference energy. With the aid of mathematical simulations and measurements on a physical phantom, the feasibility of the procedure and its insensitivity with regard to energy settings and other machine parameters are documented.

摘要

计算机断层扫描(CT)中使用的X射线的多色性使得难以建立CT值的绝对标度,并且使得患者之间的定量比较不可靠。X射线的光谱偏移取决于所测量结构内的物质分布,如果存在大量的骨、脂肪或注入的造影剂,则这种偏移是显著的。已经开发了一种材料选择性束硬化校正程序,该程序允许相对于参考能量重建线性衰减系数的良好近似值。借助于数学模拟和对物理体模的测量,记录了该程序的可行性及其对能量设置和其他机器参数的不敏感性。