Liappis N
Klin Padiatr. 1982 Jan;194(1):60-3. doi: 10.1055/s-2008-1033773.
Iron concentration was determined by the aid of the ESA Ferrochem 3050 analyzer and the bathophenanthroline method with deproteinization. Both methods showed a good precision and accuracy. The investigation was carried out in 505 sera with an iron concentration between 9 and 333 micrograms/dl. The comparison of the two methods in 360 sera with normal iron concentration (40-150 micrograms/dl) showed a close correlation (correlation coefficient r = 0.965). Close correlation (correlation coefficient r = 0.939) was also found in the 35 sera with increased iron concentration (150-333 micrograms/dl) and acceptable correlation (correlation coefficient r = 0.808) resulted in the 110 sera with decreased iron concentration (9-40 micrograms/dl). The special merits of the potential coulometric method were precision, quickness in the time of analysis and very small sample size (50 microliters), important for pediatric analyses.
铁浓度通过ESA Ferrochem 3050分析仪和经去蛋白处理的邻二氮菲法测定。两种方法均显示出良好的精密度和准确性。该研究在505份铁浓度介于9至333微克/分升之间的血清中进行。对360份铁浓度正常(40 - 150微克/分升)的血清进行的两种方法比较显示出密切相关性(相关系数r = 0.965)。在35份铁浓度升高(150 - 333微克/分升)的血清中也发现了密切相关性(相关系数r = 0.939),而在110份铁浓度降低(9 - 40微克/分升)的血清中得到了可接受的相关性(相关系数r = 0.808)。电位库仑法的特殊优点在于精密度、分析速度快以及样品量非常小(50微升),这对儿科分析很重要。