Complement activation by nylon- and polypropylene-looped prosthetic intraocular lenses.

作者信息

Tuberville A W, Galin M A, Perez H D, Banda D, Ong R, Goldstein I M

出版信息

Invest Ophthalmol Vis Sci. 1982 Jun;22(6):727-33.

PMID:7076418
Abstract

Uncomplicated placement of an intraocular lens at the time of routine cataract extraction is associated with a greater inflammatory response than simple lens extraction. We have shown that certain components of prosthetic intraocular lenses are capable of activating the complement system in human serum in such a fashion as to generate peptides capable of mediating acute inflammatory reactions. Normal human serum was incubated with intraocular lenses in vitro and then examined for cleavage products of C3 and C5 and for C5-derived chemotactic activity for human polymorphonuclear leukocytes. Polymethylmethacrylate lenses with nylon and polypropylene loops were found to be capable of causing cleavage of C3 (detected by crossed immunoelectrophoresis) and C5 (detected by single-dimension radioimmunoelectrophoresis) as well as generation of C5-derived chemotactic activity in human serum. Serum incubated with polymethylmethacrylate lenses without loop supports exhibited no such activity. Results of our studies indicate that certain nylon- and polypropylene-looped intraocular lenses are not "inert" biologically and may elicit acute inflammatory reactions.

摘要

在常规白内障摘除术时单纯植入人工晶状体比单纯晶状体摘除引发更强的炎症反应。我们已经表明,人工晶状体的某些成分能够以产生能够介导急性炎症反应的肽的方式激活人血清中的补体系统。将正常人血清与人工晶状体在体外孵育,然后检测C3和C5的裂解产物以及C5衍生的对人多形核白细胞的趋化活性。发现带有尼龙和聚丙烯袢的聚甲基丙烯酸甲酯人工晶状体能够引起C3裂解(通过交叉免疫电泳检测)和C5裂解(通过单向放射免疫电泳检测)以及在人血清中产生C5衍生的趋化活性。与没有袢支撑的聚甲基丙烯酸甲酯人工晶状体孵育的血清没有这种活性。我们的研究结果表明,某些带有尼龙和聚丙烯袢的人工晶状体在生物学上并非“惰性”,可能引发急性炎症反应。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验