Dannehl K, Berger H, Koschinsky T, Toeller M, Gries F A
Dtsch Med Wochenschr. 1982 Jul 23;107(29-30):1135-9. doi: 10.1055/s-2008-1070089.
Thirty small reflectometers not requiring mains electricity (23 Glucosemeter and 7 individual appliances of various makes) were tested under standardised laboratory conditions as to their relative and absolute accuracy in the measurement of blood glucose. The hexokinase method delivered the reference values. The investigation demonstrated systematic deviations of 1 to 97%, coincidental deviations of 8 to 24% and total deviations of between 25 and 153%. If one accepts a value of up to 50% for the total deviation only 13 out of 30 reflectometers are acceptable. The results show that each individual meter must be assessed by the doctor before it can be given to the patient for glucose monitoring at home.