Schlebusch H, Niesen M, Paffenholz I
Abteilung für Klinische Chemie, Universität Bonn.
Klin Padiatr. 1993 Mar-Apr;205(2):111-5. doi: 10.1055/s-2007-1025208.
With the reflectance meter Reflolux S (Accucheck III) high as well as low blood glucose levels can be determined with excellent precision irrespective of the sample's hematocrit value. Within the range from 20-50 mg/dl we found an average deviation from the expected value of less than 2 mg/dl; in selected samples, however, a difference of up to 10 mg/dl was recorded. The scatter can be explained by the influence of hematocrit values on the results. An increase of 10% in hematocrit lowered the measured glucose value by approximately 10%. At glucose concentrations below 40 mg/dl and in the presence of high resp. low hematocrit values any results obtained with this glucose meter will only have screening character; they should be re-run using an established method.