Beyerl D
Am J Med Technol. 1982 May;48(5):355-9.
Comparison of transcutaneous (TC) monitoring of blood oxygen levels to arterial blood gas analyses was made on patients at rest with room air, during exercise, and at rest with oxygen. Three different transcutaneous monitors were evaluated: Novametrix TC O2 Mette, Biochem Sensomat, and Radiometer TC M1. The Hewlett-Packard ear oximeter for measuring oxygen saturation was also compared to oxygen saturation values calculated on the Severinghaus slide rule. Using at least one measured PO2 as a baseline, either TC monitoring or ear oximetry were valuable tools in monitoring pulmonary function.
对处于静息状态、呼吸室内空气、运动期间以及静息状态且吸氧的患者进行了经皮(TC)血氧水平监测与动脉血气分析的比较。评估了三种不同的经皮监测仪:诺华美TC O2 Mette、生化传感仪和雷度TC M1。还将用于测量血氧饱和度的惠普耳式血氧仪与根据西格哈斯计算尺计算出的血氧饱和度值进行了比较。以至少一个测得的氧分压作为基线,经皮监测或耳式血氧测定法都是监测肺功能的有价值工具。