Because both the recommended American Thoracic Society standards and the proposed Association for the Advancement of Medical Instrumentation standards are based on performance requirements, the most reasonable test signals are those that represent actual patients. To test and qualify spirometers for compliance with a standard, a set of 24 standard waveforms have been selected from a larger set of FVC maneuvers performed by patients who have a diversity of abnormalities and demonstrate a range of test results. Common spirometric parameters are known for each waveform; therefore, the adherence to performance requirements can be readily verified. These standard waveforms are stored on various digital mediums and can be used to drive a controlled mechanical syringe for testing actual hardware and software, or they can be put into a system in digital format to test only the software.
由于美国胸科学会的推荐标准和医疗仪器促进协会提议的标准均基于性能要求,所以最合理的测试信号是那些代表实际患者的信号。为了测试肺活量计并使其符合标准,从一大组由患有各种异常疾病且测试结果呈现一定范围的患者所进行的用力肺活量(FVC)动作中,选取了一组24个标准波形。每个波形的常见肺功能参数是已知的;因此,可以很容易地验证其是否符合性能要求。这些标准波形存储在各种数字介质上,可用于驱动受控机械注射器来测试实际的硬件和软件,或者可以以数字格式输入系统仅测试软件。