Chen A, Wadsö I
J Biochem Biophys Methods. 1982 Sep;6(4):297-306. doi: 10.1016/0165-022x(82)90011-2.
A test and calibration process for microcalorimeters is described. The method has been developed with particular reference to instruments used for measurements of thermal power produced by suspensions of living cells. The process investigated is the hydrolysis of triacetin in imidazole/acetic acid buffer. The power levels are regulated by changing the buffer composition. The power will decrease slowly and very nearly linearly with time. Five test solutions, power levels 7-90 microW X ml-1, have been characterized at 37 degrees C and one of them at 25 degrees C (13 microW X ml-1). The power values for these reaction mixtures can be accurately calculated (+/- 0.5%) as a function of time during extended reaction periods, about 20 h or more.
描述了一种用于微量量热计的测试和校准过程。该方法是特别针对用于测量活细胞悬浮液产生的热功率的仪器而开发的。所研究的过程是三醋精在咪唑/乙酸缓冲液中的水解。通过改变缓冲液组成来调节功率水平。功率将随时间缓慢且几乎呈线性下降。在37℃下对五种测试溶液(功率水平为7 - 90微瓦·毫升⁻¹)进行了表征,其中一种在25℃下(13微瓦·毫升⁻¹)进行了表征。在约20小时或更长的延长反应期内,这些反应混合物的功率值可以作为时间的函数准确计算(±0.5%)。