Murray P R, Niles A C
J Clin Microbiol. 1982 Dec;16(6):1152-4. doi: 10.1128/jcm.16.6.1152-1154.1982.
We determined the effect of performing antimicrobial susceptibility tests in five different anaerobic incubation systems: GasPak jar, large GasPak jar, evacuated-gassed anaerobic jar, anaerobic chamber, and Bio-Bag. Growth of the anaerobes was equivalent in all five incubation systems. The results of testing 38 anaerobes against 11 antimicrobial agents were comparable for the anaerobic jars and anaerobic chamber. However, discordant results were observed for metronidazole and cefamandole tests when incubated in the Bio-Bag.
GasPak罐、大型GasPak罐、抽气-充气厌氧罐、厌氧箱和生物袋。在所有五种培养系统中厌氧菌的生长情况相当。对38株厌氧菌针对11种抗菌药物进行测试的结果,在厌氧罐和厌氧箱中是可比的。然而,当在生物袋中培养时,甲硝唑和头孢孟多试验出现了不一致的结果。