Fraser C G, Peake M J
Ann Clin Biochem. 1978 Sep;15(5):261-4. doi: 10.1177/000456327801500162.
An evaluation of a colorimetric kit method for the determination of inorganic phosphate (Pierce Phosphorus Auto/Stat Kit) is described. The within-batch and between-batch precisions were shown to fulfil current criteria, and recovery experiments, linearity studies, analyses of quality control materials, and studies of possible interfering substances evidenced good accuracy. Comparison of the results obtained on samples from patients with those obtained by the vanadate/molybdate continuous-flow method showed that the test method had a comparative positive bias. The use of a calibration reference serum as standard is recommended. The kit method is technically simple, requiring no protein precipitation, and analyses can be performed rapidly.
本文描述了一种用于测定无机磷酸盐的比色试剂盒方法(Pierce磷自动/静态试剂盒)的评估。批内和批间精密度均符合现行标准,回收率实验、线性研究、质量控制材料分析以及可能干扰物质的研究均证明该方法具有良好的准确性。将该方法所得结果与钒酸盐/钼酸盐连续流动法对患者样本所得结果进行比较,发现该测试方法存在相对正偏差。建议使用校准参考血清作为标准。该试剂盒方法技术简单,无需进行蛋白质沉淀,且分析可快速完成。