Gillenwater J Y
J Urol. 1981 Mar;125(3):383-4. doi: 10.1016/s0022-5347(17)55043-0.
Chemstrip-L was evaluated in 300 patients for its ability to detect leukocytes in urine. The strip reacts to esterases that are present in leukocytes and not in normal urine. When the chamber count is used as a standard Chemstrip-L has a sensitivity of 95.3 per cent and a specificity of 98 per cent. A minimal reaction time of 15 minutes was confirmed. No discrepancy could be related directly to drug interference.
对300名患者的尿液进行了Chemstrip-L检测白细胞能力的评估。该试纸条对白细胞中存在而正常尿液中不存在的酯酶起反应。以镜检计数作为标准时,Chemstrip-L的灵敏度为95.3%,特异性为98%。确认最短反应时间为15分钟。未发现任何差异与药物干扰直接相关。