Appraisal of a new rapid enzyme strip.

作者信息

Nankervis A, Humphrey R, Westwood S, White K, Alford F P

出版信息

Med J Aust. 1980 Dec 13;2(12):677-8. doi: 10.5694/j.1326-5377.1980.tb131975.x.

Abstract

The accuracy and reliability of new blood glucose measurement strips, BM Test-Glycemie 20-800, were assessed in 99 consecutive patients attending the diabetic clinic and seven subjects at home. The BM Test-Glycemie 20-800 strip technique was compared with the automated hexokinase method and the Reflotest-Glucose test strip technique employing the Boehringer Reflomat. The mean of the blood glucose values was 12.6 +/- 0.6 mmol/L for the hexokinase method and 12.9 +/- 0.6 mmol/L for the 20-800 strip technique, with a mean difference between the hexokinase and the 20-800 strip methods of 1.9 +/- 0.2 mmol/L. In the range 4 mmol/L to 12 mmol/L, the 20-800 strips were found to be extremely accurate, but for blood glucose levels less than 3.5 mmol/L and greater than 12 mmol/L, they were less reliable. Importantly, the strips were often misleading in indicating hypoglycaemia. It is concluded that the BM Test-Glycemie 20-800 strips have a practical and valuable role when used by suitably instructed staff members, or by patients in the management of diabetes, and represent a major advance over urine testing on previously available direct reading enzyme strip methods.

摘要

对99名连续就诊于糖尿病门诊的患者以及7名居家受试者,评估了新型血糖检测试纸BM Test-Glycemie 20-800的准确性和可靠性。将BM Test-Glycemie 20-800试纸技术与自动己糖激酶法以及采用勃林格Reflomat的Reflotest-葡萄糖试纸技术进行了比较。己糖激酶法测得的血糖值均值为12.6±0.6 mmol/L,20-800试纸技术测得的血糖值均值为12.9±0.6 mmol/L,己糖激酶法与20-800试纸法之间的平均差值为1.9±0.2 mmol/L。在4 mmol/L至12 mmol/L的范围内,发现20-800试纸极为准确,但对于血糖水平低于3.5 mmol/L和高于12 mmol/L的情况,其可靠性较低。重要的是,这些试纸在指示低血糖方面常常具有误导性。结论是,BM Test-Glycemie 20-800试纸在由经过适当培训的工作人员或患者用于糖尿病管理时具有实际且重要的作用,并且相对于以前可用的直接读数酶试纸法的尿液检测而言是一项重大进步。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验