五种肺活量计的临床评估。莫纳根M403、肺功能筛查仪、呼吸功能测试仪、维卡测试肺活量计和伟康图。

Clinical evaluation of five spirometers. Monaghan M403, Pneumoscreen, Spirotron, Vicatest and Vitalograph.

作者信息

Wever A M, Britton M G, Hughes D T, Van der Plas K H, Wever-Hess J

出版信息

Eur J Respir Dis. 1981;62(2):127-37.

PMID:7238663
Abstract

Five spirometers were compared with the Stead-Wells spirometer in 100 subjects, using the FVC and FEV1 as parameters. In addition, some physical characteristics were studied. The Stead-Wells is a water-sealed spirometer, recommended as a reference spirometer by the American Committees on Environmental Health and Respiratory Physiology. The Vicatest and Vitalograph are volume-displacement dry spirometers. They showed a very good linearity over their entire volume ranges; their FVC and FEV1 readings showed very good correlations with the readings of the Stead-Wells spirometer. The Monaghan, Pneumoscreen, and Spirotron are flow-integrating "electronic" spirometers. The volumes recorded were found to be dependent on the values of the flow; their FVC and FEV1 readings also showed good correlations with the readings of the Stead-Wells spirometer, but to a lesser extent than the two dry spirometers. Therefore, for routine assessment of ventilatory function we prefer, at present, the modern volume-displacement dry spirometers.

摘要

以用力肺活量(FVC)和第一秒用力呼气容积(FEV1)作为参数,将五台肺量计与100名受试者的Stead-Wells肺量计进行了比较。此外,还研究了一些身体特征。Stead-Wells是一种水封式肺量计,被美国环境卫生与呼吸生理学委员会推荐为参考肺量计。Vicatest和Vitalograph是容积置换干式肺量计。它们在整个容积范围内显示出非常好的线性;它们的FVC和FEV1读数与Stead-Wells肺量计的读数显示出非常好的相关性。Monaghan、Pneumoscreen和Spirotron是流量积分“电子”肺量计。发现记录的容积取决于流量值;它们的FVC和FEV1读数与Stead-Wells肺量计的读数也显示出良好的相关性,但程度低于两台干式肺量计。因此,就通气功能的常规评估而言,目前我们更倾向于现代容积置换干式肺量计。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验