On the hemolytic and thrombogenic potential of occluder prosthetic heart valves from in-vitro measurements.

作者信息

Figliola R S, Mueller T J

出版信息

J Biomech Eng. 1981 May;103(2):83-90. doi: 10.1115/1.3138265.

Abstract

An experimental investigation was conducted to determine the magnitude of shear stresses and areas of stasis of several types of prosthetic occluder heart valves. These experiments were performed in a steady-flow test loop using an axisymmetric aortic-shaped test chamber and an aqueous-glycerine solution. The flow loop produced a low-turbulence intensity and uniform mean velocity profile upstream of the test chamber. Tests were performed on a Kay-Shiley disk, a Bjork-Shiley tilting disk and Starr-Edwards Models 1260 and 2320 ball prostheses at Reynolds numbers between 2000 and 6200. Momentum transfer and turbulence data were obtained both around and distal to the valve occluders using laser Doppler and hot-film anemometry. The region directly surrounding the valve occluders contained the largest stresses measured. Aortic wall shear measurements revealed magnitudes potentially damaging to the vessel lining. Regions of slowly moving separated flow found to exist in these occluder valve flow fields correlated with clinical findings of thrombus formation.

摘要

进行了一项实验研究,以确定几种类型的人工瓣膜心脏瓣膜的剪应力大小和血流停滞区域。这些实验在一个稳流测试回路中进行,该回路使用一个轴对称的主动脉形状的测试腔和一种水甘油溶液。流动回路在测试腔上游产生了低湍流强度和均匀的平均速度分布。在雷诺数介于2000和6200之间时,对凯-希利盘式瓣膜、比约克-希利倾斜盘式瓣膜以及斯塔尔-爱德华兹1260型和2320型球瓣假体进行了测试。使用激光多普勒和热膜风速仪获取了瓣膜封堵器周围及远端的动量传递和湍流数据。瓣膜封堵器直接周围的区域包含测量到的最大应力。主动脉壁剪应力测量结果显示,其大小可能会对血管内膜造成损伤。在这些封堵器瓣膜流场中发现的缓慢移动的分离流区域与血栓形成的临床发现相关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验