Wright B M
Br Med J. 1978 Dec 9;2(6152):1627-8. doi: 10.1136/bmj.2.6152.1627.
A new miniature Wright peak-flow meter has been designed and produced. The meter is tubular with a spring-loaded piston and a longitudinal slot through which air escapes. Its dynamic characteristics have been carefully designed to make it respond only to peak flow and not to rate of rise. Performance tests on early instruments showed fairly close correlation with the Wright peak-flow meter but with a constant error of + 38 1/min. On later models the correlation was increased to 0.990 and the error reduced to + 3%. The mini-meter correlates as well with the standard instrument as two standard instruments correlate with each other and should prove useful clinically.
一种新型微型赖特峰流速仪已设计并生产出来。该仪器呈管状,带有一个弹簧加载活塞和一个空气从中逸出的纵向狭槽。其动态特性经过精心设计,使其仅对峰值流速做出响应,而不对上升速率做出响应。早期仪器的性能测试表明,与赖特峰流速仪的相关性相当紧密,但存在 +38升/分钟的恒定误差。在后来的型号中,相关性提高到了0.990,误差降低到了 +3%。这种微型仪器与标准仪器的相关性与两台标准仪器之间的相关性一样好,在临床上应该会证明是有用的。