Younger J B, Boots L R, Coleman C
Fertil Steril. 1978 Dec;30(6):648-53. doi: 10.1016/s0015-0282(16)43691-5.
A 1-day rapid assay for luteinizing hormone (RLH) was developed by combined alteration of incubation temperatures and immune-complex precipitation of a commercially available LH kit. The RLH assay, compared with the standard 3-day assay, had the following characteristics: standard curve slope values of --2.47 and --2.314; assay sensitivity of 5.5 and 3.8 mIU/ml of serum; average coefficients of variation for the standard curves of 13.6% and 14.1% and for serum controls of 10.6% and 13.1%, respectively. Two consecutive RLH assays of split samples were highly correlated (r = 0.9908; P less than 0.001), as were split aliquots measured by RLH or the 3-day assay (r = 0.993; P less than 0.001). When 52 samples from portions of 5 normal menstrual cycles were assayed in both systems, agreement was again high (r = 0.9871; P less than 0.001). Representative curves of both RLH and standard LH assays, together with other data, are presented to show the feasibility of applying this method for timing of artificial insemination.
通过改变市售促黄体生成素(LH)试剂盒的孵育温度并结合免疫复合物沉淀法,开发出一种为期1天的促黄体生成素快速检测法(RLH)。与标准的3天检测法相比,RLH检测法具有以下特点:标准曲线斜率值分别为-2.47和-2.314;检测灵敏度分别为血清5.5和3.8 mIU/ml;标准曲线的平均变异系数分别为13.6%和14.1%,血清对照的平均变异系数分别为10.6%和13.1%。对分割样本进行的两次连续RLH检测高度相关(r = 0.9908;P < 0.001),通过RLH或3天检测法对分割等分样本进行检测的结果也高度相关(r = 0.993;P < 0.001)。当对来自5个正常月经周期部分的52个样本在两个系统中进行检测时,一致性同样很高(r = 0.9871;P < 0.001)。给出了RLH和标准LH检测法的代表性曲线以及其他数据,以表明应用该方法进行人工授精时机选择的可行性。