Kreuzig F, Frank J
J Chromatogr. 1981 Nov 20;218:615-20. doi: 10.1016/s0021-9673(00)82086-5.
A high-performance liquid chromatographic method for the determination of D-penicillamine in plasma and urine has been developed, based on separation on a cation-exchange resin followed by an amperometric detection of the SH group of D-penicillamine oxidized on a gold electrode. The method has been automated and separations from plasma and urine take 7 and 9 min, respectively. The detection limits are 0.05 micrograms/ml in plasma and 0.2 micrograms/ml in urine, with a coefficient of variation of 2.9% (n = 10).