Renoe B W, Stewart K K, Beecher G R, Wills M R, Savory J
Clin Chem. 1980 Feb;26(2):331-4.
We describe an adaptation of automated multiple flow-injection analysis instrumentation to an analysis for albumin in serum. The bromcresol green reaction was used to test the utility of the system. The approach yielded albumin results with excellent sensitivity, no measurable carryover, a relative standard deviation of less than 1%, good correlations with published procedures, and no measurable interferences. The simplicity and flexibility of the instrumentation and its performance integrity, as indicated by the analytical results, make this a viable clinical chemical tool.
我们描述了一种将自动多流动注射分析仪器应用于血清白蛋白分析的方法。采用溴甲酚绿反应来测试该系统的实用性。该方法得到的白蛋白检测结果具有出色的灵敏度、无可测残留、相对标准偏差小于1%、与已发表方法具有良好的相关性且无可测干扰。仪器的简单性和灵活性以及分析结果所显示的性能完整性,使其成为一种可行的临床化学工具。