Olson R J
J Am Intraocul Implant Soc. 1980 Jan;6(1):16-7. doi: 10.1016/s0146-2776(80)80099-1.
One hundred and twenty lenses from 12 manufacturers were randomly selected for optical studies using an optical bench. Resolution efficiency was measured with an Air Force target, and measurements of back focal length in air were used to calculate back focal length in aqueous. There were no lenses with: less than 50% resolution efficiency, more than a 0.5-D variation from the stated power, or more than 0.25 D of toricity. Seventy-five percent of the lenses had at least 74% resolution efficiency; 93% were within 0.25 D of the stated power; 95% had 0.1 D or less toricity. Lenses of recent manufacture exhibit a marked improvement in optical quality. Optical parameters for all lenses observed in this study were within good clinical ranges.
从12家制造商生产的120片镜片中随机选取用于使用光具座进行光学研究。使用空军靶标测量分辨率效率,并利用空气中后焦距的测量值来计算在房水中的后焦距。没有镜片的分辨率效率低于50%,与标称屈光度的偏差超过0.50D,或散光超过0.25D。75%的镜片具有至少74%的分辨率效率;93%的镜片与标称屈光度的偏差在0.25D以内;95%的镜片散光为0.1D或更低。近期生产的镜片在光学质量上有显著提高。本研究中观察到的所有镜片的光学参数都在良好的临床范围内。