Olson R J, Waters S W
Arch Ophthalmol. 1980 Nov;98(11):2060-1. doi: 10.1001/archopht.1980.01020040912026.
One hundred forty-six lenses from 11 manufactures were tested with the Veri-Vu lensometer and optical power plus resolution compared with results from an optical bench. Accuracy was found to be +/- 0.2 diopter in 95% of the lenses tested. The greatest error was 0.37 D. The lensometer correctly identified 125 of 128 lenses that were greater than 160 line pairs per millimeter (LP/mm) resolution and correctly identified nine lenses that had less than 100-LP/mm resolution. All other important optical aberrations were identified with the lensometer. The greatest problem was the variation from company to company in terms of the calculated power in aqueous as determined from the back focal length in air. This, however, was felt to be an asset for those doing A-scan intraocular lens power determination in giving one simple reliable power for all lens companies.
对来自11家制造商的146片镜片使用Veri-Vu镜片计进行测试,并将其屈光力和分辨率与光具座的结果进行比较。在95%的测试镜片中,发现精度为±0.2屈光度。最大误差为0.37D。镜片计正确识别出128片中分辨率大于160线对/毫米(LP/mm)的125片镜片,并正确识别出9片中分辨率小于100-LP/mm的镜片。镜片计还识别出了所有其他重要的光学像差。最大的问题是不同公司之间根据空气中的后焦距计算出的房水中的屈光力存在差异。然而,对于那些进行A超人工晶状体屈光力测定的人来说,这被认为是一项优势,因为它能为所有镜片公司提供一个简单可靠的屈光力数据。