使用和不使用荧光素的压平式眼压测量法。

Applanation tonometry with and without fluorescein.

作者信息

Roper D L

出版信息

Am J Ophthalmol. 1980 Nov;90(5):668-71. doi: 10.1016/s0002-9394(14)75134-7.

Abstract

Some investigators have suggested that it is not necessary to be fluorescein when doing applanation tonometry. This is incorrect because, in order to visualize the apex of the tear order to visualize the apex of the tear meniscus which defines the applanated area, the tears must be stained with fluorescein. I found an average pressure of 11.72 +/- 3.20 mm Hg in eyes tested without fluorescein compared with an average pressure of 17.34 +/- 4.07 mm Hg when pressures were immediately remeasured with fluorescein. Failure to use fluorescein during applanation tonometry will result in an under-estimation of approximately 5.62 mm Hg.

摘要

一些研究者认为,在进行压平眼压测量时不一定非要使用荧光素。这种观点是错误的,因为为了看清界定压平区域的泪液半月板顶端,泪液必须用荧光素染色。我发现,未使用荧光素进行眼压测试的眼睛平均眼压为11.72±3.20毫米汞柱,而使用荧光素立即重新测量时平均眼压为17.34±4.07毫米汞柱。在压平眼压测量过程中不使用荧光素会导致眼压低估约5.62毫米汞柱。