Hesselson A B, Parsonnet V
Department of Surgery, University of Medicine and Dentistry of New Jersey, Newark.
Pacing Clin Electrophysiol. 1994 Aug;17(8):1332-43. doi: 10.1111/j.1540-8159.1994.tb02453.x.
Advancements in technology have made paced ECG interpretation increasingly difficult. A new method for depicting the complex pacemaker/heart interactions that eliminates the extensive use of symbols and repetitious use of refractory period and rate limit information of previous methods has been devised. The method uses a framework of parallel horizontal lines drawn on grid paper underneath the ECG. The lines are spaced apart by the actual programmed values (lower rate, AV, VA intervals) of the pacemaker in question. This framework allows the simultaneous use of the horizontal and vertical directions for the diagram of pacemaker timing intervals. Also, a single representation of refractory periods, upper rate intervals, and other variables can be labeled vertically and extrapolated horizontally across the entire diagram. Single chamber, dual chamber, and rate-modulated ECGs are readily represented. The diagram is easily plotted on standard ECG paper and flexible enough to represent complex ECGs.
技术的进步使得有起搏心电图的解读越来越困难。一种描绘复杂起搏器/心脏相互作用的新方法已经设计出来,该方法消除了先前方法中广泛使用的符号以及对不应期和心率限制信息的重复使用。该方法使用在心电图下方的方格纸上绘制的平行水平线框架。这些线之间的间距由所讨论起搏器的实际编程值(下限频率、房室间期、室房间期)决定。这个框架允许在起搏器计时间期图中同时使用水平和垂直方向。此外,不应期、上限频率间期和其他变量的单一表示可以垂直标注,并在整个图中水平外推。单腔、双腔和频率适应性心电图都能很容易地表示出来。该图很容易绘制在标准心电图纸上,并且足够灵活以表示复杂的心电图。