Durability of laser fibers.

作者信息

te Slaa E, van Ettergem A F, van 't Hof C A, Debruyne F M, de la Rosette J J

机构信息

Department of Urology, University Hospital Nijmegen, The Netherlands.

出版信息

World J Urol. 1995;13(2):83-7. doi: 10.1007/BF00183619.

Abstract

In 90 patients treated with laser prostatectomy using the Urolase (n = 50) or Ultraline (n = 40) laser fiber, the fiber-tip durability was investigated. In general the Urolase fiber tips were less damaged than the Ultraline fiber tips. At visual inspection, 62% of the Urolase fiber tips were graded as minimally damaged in comparison with 28% of the Ultraline group. The Urolase fiber tips are more susceptible than the Ultraline fiber tips to damage caused by tissue contact, whereas the latter seem more fragile. Transmission measurements were performed in a laboratory setting to estimate the loss of energy output at the fiber tip due to damage. These measurements showed a major loss in almost all fibers. None of the Ultraline fibers had less than 10% transmission loss, and 18% of the Urolaser fibers had a transmission value of more than 90%. Finally, there seemed to be a poor correlation between the visual aspects of the fibers used and the changes in transmission.

摘要

在90例接受使用Urolase激光光纤(n = 50)或Ultraline激光光纤(n = 40)进行激光前列腺切除术的患者中,对光纤尖端的耐用性进行了研究。总体而言,Urolase光纤尖端比Ultraline光纤尖端受损程度轻。目视检查时,62%的Urolase光纤尖端被评为轻度受损,而Ultraline组为28%。Urolase光纤尖端比Ultraline光纤尖端更容易因组织接触而受损,而后者似乎更脆弱。在实验室环境中进行了传输测量,以估计由于损伤导致的光纤尖端能量输出损失。这些测量结果显示几乎所有光纤都有较大损失。没有一根Ultraline光纤的传输损失小于10%,18%的Urolaser光纤的传输值超过90%。最后,所用光纤的外观与传输变化之间似乎相关性较差。