Qian K X, Wang S S, Chu S H
Department of Surgery, Medical College of National Taiwan University, Taipei, Republic of China.
Artif Organs. 1995 Apr;19(4):328-33. doi: 10.1111/j.1525-1594.1995.tb02336.x.
A pulsatile impeller assist heart and a total heart were tested as a chronic left ventricular assist device in 5 calves and an acute biventricular assist device in 4 pigs respectively, to evaluate their blood compatibility. During the left ventricular assist experiments, the indicators for hemolysis, thrombogenesis, renal dysfunction, and hepatic dysfunction were measured preoperatively, at the beginning of the pumping, 6 h postoperatively, and every following day. The results demonstrated that the impeller assist heart causes no severe blood damage nor organ dysfunction in the experiments lasting up to 11 days. In biventricular assist experiments, the number of red blood cells, white blood cells, platelets, and the hematocrit, hemoglobin, free hemoglobin, and lactate dehydrogenase levels were tested preoperatively at the beginning of the pumping and every 2 h postoperatively. The data remained in acceptable ranges during experiments lasting 6 h. It is confirmed that the authors' impeller assist heart and total heart have the advantages of simplicity, implantability, and pulsatility with good blood compatibility.
一种搏动性叶轮辅助心脏和一种全人工心脏分别作为慢性左心室辅助装置在5头小牛身上进行了测试,并作为急性双心室辅助装置在4头猪身上进行了测试,以评估它们的血液相容性。在左心室辅助实验中,在术前、开始泵血时、术后6小时以及随后的每一天测量溶血、血栓形成、肾功能障碍和肝功能障碍的指标。结果表明,在长达11天的实验中,叶轮辅助心脏未造成严重的血液损伤或器官功能障碍。在双心室辅助实验中,术前、开始泵血时以及术后每2小时检测红细胞、白细胞、血小板数量以及血细胞比容、血红蛋白、游离血红蛋白和乳酸脱氢酶水平。在持续6小时的实验中,数据保持在可接受范围内。证实了作者的叶轮辅助心脏和全人工心脏具有结构简单、可植入性强、搏动性好以及血液相容性良好的优点。