Suppr超能文献

尖顶距离在预测髋部转子周围骨折固定失败中的价值。

The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip.

作者信息

Baumgaertner M R, Curtin S L, Lindskog D M, Keggi J M

机构信息

Department of Orthopaedics and Rehabilitation, Yale University School of Medicine, New Haven, Connecticut 06520-8071, USA.

出版信息

J Bone Joint Surg Am. 1995 Jul;77(7):1058-64. doi: 10.2106/00004623-199507000-00012.

Abstract

Failure of fixation of peritrochanteric fractures that have been treated with a fixed-angle sliding hip-screw device is frequently related to the position of the lag screw in the femoral head. A simple measurement has been developed to describe the position of the screw. This measurement, the tip-apex distance, is the sum of the distance from the tip of the lag screw to the apex of the femoral head on an anteroposterior radiograph and this distance on a lateral radiograph, after controlling for magnification. To determine the value of this measurement in the prediction of so-called cutout of the lag screw, 198 peritrochanteric fractures (193 patients) were studied. The minimum duration of follow-up was three months (average, thirteen months), during which period all of the fractures either healed or had failure of the fixation. Of the nineteen failures that were identified, sixteen were due to the device cutting out of the femoral head. The average tip-apex distance was twenty-four millimeters (range, nine to sixty-three millimeters) for the successfully treated fractures compared with thirty-eight millimeters (range, twenty-eight to forty-eight millimeters) for those in which the screw cut out (p = 0.0001). None of the 120 screws with a tip-apex distance of twenty-five millimeters or less cut out, but there was a very strong statistical relationship between an increasing tip-apex distance and the rate of cutout, regardless of all other variables related to the fracture.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

采用角钢板滑动髋螺钉装置治疗的转子周围骨折固定失败,常常与拉力螺钉在股骨头内的位置有关。现已开发出一种简单的测量方法来描述螺钉的位置。这种测量方法,即尖顶距,是在前后位X线片上拉力螺钉尖端至股骨头顶点的距离与侧位X线片上该距离(校正放大率后)之和。为了确定该测量方法在预测拉力螺钉所谓穿出方面的价值,对198例转子周围骨折(193例患者)进行了研究。随访最短时间为3个月(平均13个月),在此期间所有骨折要么愈合,要么固定失败。在确定的19例固定失败病例中,16例是由于装置穿出股骨头所致。成功治疗的骨折,平均尖顶距为24毫米(范围9至63毫米),而螺钉穿出的骨折,平均尖顶距为38毫米(范围28至48毫米)(p = 0.0001)。尖顶距为25毫米或更小的120枚螺钉无一穿出,但无论与骨折相关的所有其他变量如何,尖顶距增加与穿出率之间存在非常强的统计学关系。(摘要截短至250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验