使用过氧乙酸进行液体化学消毒。一种内镜处理的替代方法。

Liquid chemical sterilization using peracetic acid. An alternative approach to endoscope processing.

作者信息

Wallace C G, Agee P M, Demicco D D

机构信息

Community Hospital of Roanoke Valley, VA 24029, USA.

出版信息

ASAIO J. 1995 Apr-Jun;41(2):151-4.

PMID:7640418
Abstract

Recurrent episodes of endoscope contamination with nontuberculous mycobacterium and Pseudomonas species, coupled with employee concerns about exposure to 2% glutaraldehyde and the requirement for rapid scope turnover time, led to the investigation of an alternative method for endoscope processing. A prospective evaluation of 220 bronchoscopy procedures was carried out. Endoscope culture surveillance was performed twice a month for a 12 month period on all endoscopes. In addition, the deliberate inoculation of bronchoscopes with Mycobacterium gordonae was carried out. No indications of cross contamination were noted among the 220 bronchoscopy procedures after processing the scopes by the studied liquid chemical sterilization system using peracetic acid. Ongoing patient surveillance did not reveal any indications of cross contamination for nearly 15,000 patients diagnostic or surgical endoscopic procedures using instruments processed by this liquid chemical sterilization system. Twice a month for 12 months, bronchoscopes, colonoscopes, and gastroscopes were cultured for bacterium, fungi, and acid fast bacilli and showed no growth. Deliberate inoculation of bronchoscopes with M. gordonae showed the scopes were rendered sterile. No unusual occurrences of occupational chemical exposure were found in 11 individuals working with this liquid chemical sterilization system. The peracetic acid based sterilization system studied proved to be an effective, safe, and economical alternative to the glutaraldehyde method.

摘要

非结核分枝杆菌和假单胞菌属反复污染内窥镜,再加上员工对接触2%戊二醛的担忧以及内窥镜快速周转时间的要求,促使人们对一种替代的内窥镜处理方法进行研究。对220例支气管镜检查程序进行了前瞻性评估。在12个月的时间里,每月对所有内窥镜进行两次培养监测。此外,还故意用戈登分枝杆菌接种支气管镜。在用过氧乙酸的液体化学消毒系统处理内窥镜后,220例支气管镜检查程序中未发现交叉污染迹象。对近15000例使用该液体化学消毒系统处理器械进行诊断或手术内窥镜检查的患者进行的持续监测未发现任何交叉污染迹象。在12个月的时间里,每月两次对支气管镜、结肠镜和胃镜进行细菌、真菌和抗酸杆菌培养,均未生长。用戈登分枝杆菌故意接种支气管镜表明这些镜子已无菌。在使用该液体化学消毒系统的11名工作人员中未发现异常的职业化学暴露情况。所研究的基于过氧乙酸的消毒系统被证明是一种有效、安全且经济的替代戊二醛方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验