Hunter J S
Department of Pathology and Laboratory Medicine, Christ Hospital and Medical Center, Oak Lawn, Illinois 60453.
J Clin Microbiol. 1993 Feb;31(2):465-6. doi: 10.1128/jcm.31.2.465-466.1993.
Despite the customization of growth index thresholds within individual laboratories, use of the BACTEC NR 660 automated blood culture system results in a number of false-positive cultures. The results of Gram staining, acridine orange staining, and subculturing to agar media were evaluated on 210 false-positive blood cultures over a 6-month period. Inclusion of acridine orange staining in the routine workup of false-positive blood cultures can eliminate the need for subculturing.
尽管各个实验室对生长指数阈值进行了定制,但使用BACTEC NR 660自动血培养系统仍会产生一些假阳性培养结果。在6个月的时间里,对210份假阳性血培养物进行了革兰氏染色、吖啶橙染色及转种至琼脂培养基的结果评估。在假阳性血培养物的常规检查中加入吖啶橙染色可无需进行转种培养。