Greenhut S E, Svinarich J T, Randall N J, Heald S C, Nappholz T A
Applied Research Division, Telectronics Pacing Systems, Englewood, Colorado 80112.
Pacing Clin Electrophysiol. 1993 Jun;16(6):1293-303. doi: 10.1111/j.1540-8159.1993.tb01716.x.
The detection of atrial activation from a standard ventricular pacing lead with standard ventricular electrodes would provide patients with VVI and VVIR pacing systems atrial rate response and atrial synchrony. In addition to potentially increasing cardiac output appropriately in these patients at rest and during moderate exercise, P wave sensing with such a device could help reduce pacemaker syndrome. In this study, unipolar signals from distal and proximal intraventricular electrodes were recorded from the right ventricular apex in 20 patients. Unipolar electrograms from 16 patients were recorded using temporary electrophysiology catheters and in four patients using permanent pacemaker leads. Approximately 3 minutes of data per patient were acquired and analyzed. After selection of a P wave template, the difference in baseline normalized area between the template and signal was calculated on a point-by-point basis. The percent of atrial depolarizations correctly detected was determined for each patient and lead configuration at the optimal threshold. Far-field P wave accuracy was better at the proximal electrode (74 +/- 25%) than at the distal electrode (57 +/- 34%) (P < 0.025). At the proximal electrode, 15/20 (75%) patients had > 70% accuracy and 11/20 (55%) patients had > 80% accuracy. At the distal electrode, 10/21 (48%) patients had > 70% accuracy and 7/21 (33%) patients had > 80% accuracy. In conclusion, far-field detection of atrial activation at the ventricular proximal electrode appears possible with sufficient accuracy to provide periods of atrial rate response and synchrony in patients with a single standard lead.
利用标准心室电极从标准心室起搏导线检测心房激动可为采用VVI和VVIR起搏系统的患者提供心房率应答和心房同步功能。除了可能在这些患者静息和中度运动时适当增加心输出量外,使用这种装置进行P波感知有助于减少起搏器综合征。在本研究中,从20例患者的右心室心尖记录了来自远端和近端心室内电极的单极信号。16例患者使用临时电生理导管记录单极电图,4例患者使用永久起搏器导线记录。每位患者采集并分析约3分钟的数据。选择P波模板后,逐点计算模板与信号之间基线归一化面积的差异。在最佳阈值下,确定每位患者和导联配置正确检测到的心房去极化百分比。近端电极处的远场P波准确性(74±25%)优于远端电极处(57±34%)(P<0.025)。在近端电极,15/20(75%)的患者准确性>70%,11/20(55%)的患者准确性>80%。在远端电极,10/21(48%)的患者准确性>70%,7/21(33%)的患者准确性>80%。总之,在心室近端电极处进行心房激动的远场检测似乎具有足够的准确性,可为采用单根标准导线的患者提供心房率应答和同步期。