Placement of umbilical venous catheters with use of bedside real-time ultrasonography.

作者信息

Greenberg M, Movahed H, Peterson B, Bejar R

机构信息

Department of Neonatology, University of California-San Diego, USA.

出版信息

J Pediatr. 1995 Apr;126(4):633-5. doi: 10.1016/s0022-3476(95)70366-7.

Abstract

Real-time ultrasonography was used to assess the position of umbilical venous catheters tips in 79 newborn infants (birth weight range, 400 to 4200 gm). The position of the umbilical venous catheters determined by ultrasonography was compared with the projection of the catheter tip on the closest vertebral body shown on a standard chest radiograph. Of the catheters initially placed with radiographic guidance, 56% were repositioned because of unsatisfactory location. Ultrasonography is more precise than radiography because it allows direct visualization of the catheter tips in relation to internal vascular structures. When ultrasonography is unavailable, placement at the ninth thoracic vertebral body ensures safe positioning.

摘要

采用实时超声检查评估了79例新生儿(出生体重范围为400至4200克)脐静脉导管尖端的位置。将超声检查确定的脐静脉导管位置与标准胸部X线片上显示的导管尖端在最接近椎体上的投影进行比较。在最初在X线引导下放置的导管中,56%因位置不理想而重新定位。超声检查比X线检查更精确,因为它能直接显示导管尖端与内部血管结构的关系。当无法进行超声检查时,放置在第9胸椎椎体可确保安全定位。