使用简易光度计对医学成像用阴极射线管显示器进行质量控制。

Quality control of cathode-ray tube monitors for medical imaging using a simple photometer.

作者信息

Parsons D M, Kim Y, Haynor D R

机构信息

Department of Electrical Engineering, University of Washington, Seattle 98195, USA.

出版信息

J Digit Imaging. 1995 Feb;8(1):10-20. doi: 10.1007/BF03168051.

Abstract

As computer monitors are used more in medical imaging and the use of picture archiving and communication system workstations with multiple monitors is increasing, quality control protocols become necessary to track subtle variations in performance characteristics. Several tests based on previously published work were applied to 10 monitors of three different types over a period of 5 months. Each test is explained, and the results are shown. For example, without corrective adjustments, 2 monitors from the same workstation showed a small but steady decline in maximum luminance of 8.1% and 7.6% over the course of 11 weeks that was not perceptible. From this experience, we took the first step toward developing a practical and useful quality control protocol. The proposed protocol requires only a photometer and the ability to generate the Society of Motion Picture and Television Engineers (SMPTE) test pattern, and takes approximately 5 minutes per monitor per week to gather data.

摘要

随着计算机显示器在医学成像中的使用越来越多,以及配备多台显示器的图像存档与通信系统工作站的使用不断增加,质量控制协议对于跟踪性能特征的细微变化变得十分必要。在5个月的时间里,我们对三种不同类型的10台显示器应用了基于先前发表作品的多项测试。每项测试都进行了说明,并展示了测试结果。例如,在未经校正调整的情况下,同一工作站的2台显示器在11周内最大亮度出现了虽小但稳定的下降,降幅分别为8.1%和7.6%,但这种下降并不明显。基于这一经验,我们朝着制定实用的质量控制协议迈出了第一步。所提议的协议仅需要一台光度计以及生成美国电影电视工程师协会(SMPTE)测试图案的能力,并且每周每台显示器收集数据大约需要5分钟。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验