Intra-arterial pressure measurement in neonates: dynamic response requirements.

作者信息

van Genderingen H R, Gevers M, Hack W W

机构信息

Department of Clinical Physics and Engineering, Free University Hospital, Amsterdam, The Netherlands.

出版信息

Physiol Meas. 1995 Feb;16(1):55-61. doi: 10.1088/0967-3334/16/1/007.

Abstract

A computer simulation of a catheter manometer system was used to quantify measurement errors in neonatal blood pressure parameters. Accurate intra-arterial pressure recordings of 21 critically ill newborns were fed into this simulated system. The dynamic characteristics, natural frequency and damping coefficient, were varied from 2.5 to 60 Hz and from 0.1 to 1.4, respectively. As a result, errors in systolic, diastolic and pulse arterial pressure were obtained as a function of natural frequency and damping coefficient. Iso-error curves for 2%, 5% and 10% were constructed. Using these curves, the maximum inaccuracy of any neonatal catheter manometer system can be determined and used in the clinical setting.

摘要

利用导管测压系统的计算机模拟来量化新生儿血压参数的测量误差。将21名危重新生儿的准确动脉内压记录输入到这个模拟系统中。动态特性、固有频率和阻尼系数分别在2.5至60赫兹以及0.1至1.4之间变化。结果,获得了收缩压、舒张压和脉压的误差作为固有频率和阻尼系数的函数。构建了2%、5%和10%的等误差曲线。利用这些曲线,可以确定任何新生儿导管测压系统的最大不准确性,并应用于临床环境。